Decision of the Council of the Eurasian Economic Commission No. 27 of February 12, 2016: Requirements for Medical Devices

March 1, 2026
Decision No. 27 of the Council of the Eurasian Economic Commission of February 12, 2016 sets uniform requirements for medical devices placed on the market of the Eurasian Economic Union. The document defines the basic approaches to the safety, quality, design, labeling and conformity assessment of medical devices.
The requirements of Decision No. 27 apply to state registration of medical devices within the EAEU and form the basis for assessing their compliance with regulatory standards. In effect, the document sets the overall technical and regulatory framework on which registration dossiers, testing and post-market control of products rely.

Below is a structured overview of the key provisions of Decision No. 27 without excessive legal jargon. It explains the logic of the requirements and their practical significance for manufacturers and applicants working on medical device registration in the EAEU.

General provisions and key definitions

Decision No. 27 sets the general rules by which the safety and efficacy of medical devices are assessed in the EAEU. It is used during registration and determines which characteristics of a device are acceptable in terms of risk, intended use and conditions of use.

The requirements are built around intended use and the risks associated with use. The assessment is not limited to formal compliance with parameters: design, materials, operating conditions and the information provided to the user are all taken into account.

Medical devices and in vitro diagnostic devices

Medical devices include apparatus, equipment, materials and software used for treatment and diagnosis. Standalone software is treated as an active device and is subject to the general safety and efficacy requirements.

For in vitro diagnostics, requirements are set for measurement accuracy, handling of biological samples and interpretation of results.

Safety, efficacy and risk

The safety of a medical device is assessed in terms of possible harm to the user and third parties when the device is used as intended. A complete absence of risk is not expected; what is assessed is whether the residual risk is acceptable given the expected benefit.

Efficacy means that the device fulfils its stated intended use. It is confirmed by test results, clinical data and justification of the technical solutions chosen by the manufacturer.

Risks are identified and analyzed at the design stage. The manufacturer must reduce them through design and process measures and disclose information about residual risks in the labeling and instructions.

Essential safety and efficacy requirements

Decision No. 27 is used when registering medical devices in the EAEU. It is the basis for checking whether a device is safe in normal operation and fit for the intended use stated by the manufacturer.

A device must work properly under the conditions it was designed for and must not create unnecessary risks. Anything that could go wrong is either addressed in the design or described explicitly in the labeling and instructions.

Chemical, physical and biological properties

Materials are chosen based on where and how the device is used. Their behavior in contact with tissues, fluids and other media the device encounters during operation is examined.
Specific aspects are checked:

  • toxicity and biocompatibility;
  • flammability and temperature resistance;
  • wear and mechanical strength;
  • possible release of harmful substances;
  • the effect of contaminants and manufacturing residues.
If the device is used together with medicinal products, their compatibility is assessed. The device and the medicinal product must not interfere with each other's intended performance.

Control of infectious and microbial contamination

Infection safety requirements cover both the design of medical devices and the conditions of their manufacture, packaging and storage. The main goal is to reduce the risk of infection during use, maintenance and contact with the product.

The design must be clear and predictable in operation. This reduces the likelihood of microbial contamination during use and maintenance, as well as the risk of transferring contamination to the patient or the environment.

For sterile medical devices, the requirements focus on maintaining sterility until the package is opened. Sterilization and processing must be validated and reproducible, and storage conditions must match the stated parameters.

Non-sterile devices are packaged so that they remain clean and intact. If sterilization before use is intended, the packaging must be compatible with the specified method and must not affect the device's properties.

Devices containing medicinal substances and biological materials

If a device contains a medicinal substance, the substance is assessed not on its own but together with the product. The assessment looks at how it works in this specific form and whether its properties change when used together.

When biological materials are used, their origin and processing are important. For materials of animal origin, veterinary control and the health of the animals are taken into account. For materials of human origin, donor selection, storage and processing conditions are considered. This information must be documented and traceable.

The risk of transmitting infections is checked separately. The processing and inactivation methods used must reduce these risks. If destroying the activity of the biological material would make the device unusable, the requirements are applied with regard to the intended use and real conditions of use.

Protection against external factors and radiation

The design and operation take into account the conditions in which the equipment is actually used. The requirements aim to keep it safe and stable under normal exposure.

The assessment covers the following risks:

  1. mechanical damage and injury during operation and maintenance;
  2. user errors related to the design and interface;
  3. the effect of temperature, humidity, pressure and other environmental conditions;
  4. exposure to electromagnetic fields and electrostatic discharge;
  5. operation alongside other equipment without mutual interference;
  6. ignition or explosion when flammable substances are used.
Separate requirements apply to radiation sources. Radiation levels must be kept to the minimum needed to achieve the medical purpose. For potentially hazardous radiation, monitoring and warning means are provided, and the parameters must be reproducible within set limits.

Medical devices with software and active devices

If software is part of a device, it is treated as a full component of the medical product. This applies both to embedded software and to standalone software. Stability, predictable behavior and the absence of failures that could affect safety or the outcome of use are checked.

For active medical devices, risks related to energy sources are assessed separately. A single fault must not create a hazardous situation. If safety depends on the power supply, means of indicating its status and fault alarms are provided.

Active devices must not interfere with other equipment and must at the same time be resistant to external interference. Protection against electric shock during normal operation and in the event of failure is also checked.

Labeling and operational documentation

Labeling makes it immediately clear what the product is, who made it and how to use it. It is used to identify the product and to distinguish versions, batches and conditions of use.

The labeling shows basic information:

  • name and intended use;
  • manufacturer details and country of origin;
  • batch or serial number;
  • safe use period or year of manufacture;
  • storage and handling conditions;
  • indications of sterility, single use and special restrictions.
If the device contains medicinal substances, biological materials or potentially hazardous components, this is also shown on the labeling. Registered products bear the EAEU market circulation mark where technically feasible and where it creates no risk during use.

Operational documentation explains exactly how to use the product. The instructions state the intended use, main characteristics, possible risks, and installation, maintenance and storage procedures. The format may be paper or electronic; what matters is that the user has access to the current version.

For low-risk products, shortened instructions are allowed if safe use is clear without detailed explanation.

Requirements for in vitro diagnostic medical devices

Separate requirements apply to products used in laboratory diagnostics. They relate to handling biological samples, measurement accuracy and risks that arise not during treatment but when results are obtained and interpreted.

For this category, the reliability of results is essential. Measurement inaccuracies change how data are interpreted and what clinical decisions are made, so the requirements are stricter.

Performance characteristics and safety

The assessment considers how stably and predictably the diagnostic solution performs under the stated conditions. It checks whether the characteristics are maintained throughout the period of use and whether results are distorted by materials, environmental conditions or interaction with the sample.

Analytical and clinical performance indicators are particularly important. As part of confirming performance, the following are reviewed:

  • accuracy, repeatability and reproducibility of measurements;
  • sensitivity, specificity and limit of detection;
  • the effect of interfering substances and cross-reactions.
If the result depends on calibrators or control materials, their traceability must be ensured. Numerical values are expressed in standardized units so that results can be compared and interpreted correctly.

Infection safety and biological materials

When working with biological samples, the risk of infecting the user and others is taken into account. The design and packaging must minimize the likelihood of contamination during use, storage and transport.

When materials of animal, human or microbial origin are used, their origin and processing are checked. Processing methods are chosen to reduce infection risks without distorting diagnostic results. If inactivating the biological material would make the product unusable, the requirements are applied with regard to the intended use and real conditions of use.

Effect of external factors and operation as part of systems

Diagnostic solutions must work correctly under real operating conditions. This applies both to the laboratory environment and to operation together with other equipment and software.

The assessment takes into account:

  • exposure to temperature, humidity and electromagnetic fields;
  • resistance to interference when operating with other equipment;
  • risks of user error and incorrect sample identification;
  • safety in the event of single faults and power failures.
Active solutions and software components are checked separately. They must operate stably, cause no interference and not distort results in normal operation.

Labeling and instructions for use

Labeling and instructions for in vitro diagnostics contain more information than the general requirements call for. This is because test conditions and the interpretation of results must be understood precisely.

The documentation states:

  • the purpose of the test, the sample type and the principle of analysis;
  • storage conditions, reagent stability and sample preparation;
  • the test procedure and rules for interpreting results;
  • warnings, limitations and disposal requirements.
For self-testing solutions, the documentation also describes what the user should do in the event of a positive, negative or inconclusive result and states that medical decisions must not be made without consulting a specialist.

Conformity assessment procedures and registration of medical devices

During registration, Decision No. 27 serves as the basis for checking product safety and efficacy. It is used to check whether risks have been addressed and whether the stated characteristics have been confirmed.

The manufacturer can take one of two routes: either demonstrate compliance with the requirements of the document directly or rely on standards whose application is recognized as an acceptable way of demonstrating conformity. Both options are equally valid and are often combined.

Form for presenting information and how to complete it

When applying for registration, a document is prepared that states, for each requirement, whether it applies to the specific case and how compliance has been confirmed. This is not a formality but a working table that links the requirements of Decision No. 27 to actual evidence.

It records:

  • which clauses apply;
  • how exactly compliance has been confirmed;
  • which documents confirm it;
  • what conclusion has been reached for each requirement.
Clinical justification is mandatory for key safety and efficacy issues. Statements "by description" are not enough here; data and study results are needed.

Standards and voluntary application

Instead of detailed evidence for each clause, standards from the approved list may be used. These standards are updated and applied on a voluntary basis, but in practice they greatly simplify work on the dossier.

Registration projects typically use:

  • standards that fully cover individual requirements;
  • tests under standardized methods;
  • results of tests performed by third parties;
  • a combination of several approaches if the product is complex.

Conclusion

EEC Council Decision No. 27 is used to check how ready a product is for the EAEU market overall: in terms of risks, stated intended use and how it behaves in real operation.

The document sets out the general logic of the assessment, from materials and design to software, labeling and the body of evidence. For in vitro diagnostics, this logic is supplemented by requirements for accuracy, stability of results and handling of biological samples. At registration, it all comes together: the requirements are met either directly or through standards, but always with a clear justification.

The practical value of Decision No. 27 is that it reveals a product's weak points in advance. The earlier the requirements are taken into account during development and dossier preparation, the fewer questions arise at the registration and control stages.

Decision No. 27 checklist for EAEU registration

Before applying for registration, check the following:|

  • the intended use and scope of application are described clearly and do not contradict the design and characteristics;
  • risks have been identified in advance and reduced through design, and residual risks are reflected in the documentation;
  • materials and components are compatible with the media and conditions of use;
  • infection risks have been addressed, and processing and sterilization processes are reproducible;
  • software and active elements operate stably and predictably;
  • the labeling contains all mandatory information and does not mislead the user;
  • the instructions match real use scenarios and the product version;
  • for in vitro diagnostics, analytical and clinical performance has been confirmed;
  • a clear method of demonstrating conformity has been chosen, either directly or through standards;
  • clinical justification has been prepared where it is actually required.